Page | 10 INSTALLATION (CONTINUED) INSTALL SIDE AIR DUCTS SRI20PF Two (2) side air ducts are packed with the accessories of the SRI20PF. Install t
11 | P a g e INSTALLATION (CONTINUED) SHELVING INSTALLATION SRI6PF and SRI20PF Standard Shelving Installation Perform the following steps to inst
Page | 12 INSTALLATION (CONTINUED) SRI6PF and SRI20PF Sliding Shelf Installation Note: The SRI6PF does not come with a sliding shelf. Sliding shel
13 | P a g e INSTALLATION (CONTINUED) 4. Repeat the process on the opposite side of the chamber with the second sliding mounting bracket. 5. Ha
Page | 14 CONTROL PANEL OVERVIEW Figure 9: SRI6PF and SRI20PF Control Panel Main Display The main display consists of 4 operational elements. Th
15 | P a g e CONTROL PANEL (CONTINUED) Main Temperature and Timer Control The control consists of an UP/DOWN arrow pad. It is used to enter temper
Page | 16 OPERATION THEORY OF OPERATION Each SRIPF incubator uses a Peltier thermoelectric cooling-and-heating device and an air circulating fan in
17 | P a g e OPERATION (CONTINUED) SETTING UP THE INCUBATOR Perform the following steps each time the incubator is installed in a new location: 1
Page | 18 OPERATION (CONTINUED) Figure 10: Details of the SRI6PF and SRI20PF Controls and Displays SET THE CURRENT TIME 1. To change the clock to
19 | P a g e OPERATION (CONTINUED) SET THE NIGHT PHASE START POINT 1. The Temperature Display will remain dark. The Day Time indicator LED will e
Page | 2 Sheldon Fly Refrigerated Incubators 100 - 120 Voltage Peltier Installation and Operation Manual Part Number (Manual): 4861674-1 Revision:
Page | 20 OPERATION (CONTINUED) SET THE NIGHT TIME PHASE TEMPERATURE SET POINT 1. Select the Night Temp Mode by pressing and holding the Mode but
21 | P a g e OPERATION (CONTINUED) CALIBRATE THE TEMPERATURE DISPLAY Allow the incubator at least 24 hours to stabilize at its operational set poin
Page | 22 OPERATION (CONTINUED) Part 2: Calibrate the Night Time Temperature Note: The SRI6PF and SRI20PF must be calibrated to separate Day Time
23 | P a g e OPERATION (CONTINUED) SET THE OVER TEMPERATURE LIMIT CONTROL Perform the following steps to set up the Over Temperature Limit system
Page | 24 OPERATION (CONTINUED) SILENCING THE DOOR AJAR ALARM If the door alarm needs to be silenced, close the door, which will reset the timer. Fo
25 | P a g e USER MAINTENANCE Warning: Prior to any maintenance or service on this unit, disconnect the power cord from the power supply. Avertis
Page | 26 MAINTENANCE (CONTINUED) Periodically, inspect the door latch, trim, catch, and gasket for signs of deterioration. Failure to maintain the
27 | P a g e UNIT SPECIFICATIONS These incubators are 100 - 120 voltage units. Please refer to each incubator’s data plate for its electrical speci
Page | 28 UNIT SPECIFICATIONS (CONTINUED) SHELF CAPACITY BY WEIGHT Model Per Shelf Total SRI6PF 75lbs / 34kg 150lbs / 34kg SRI20PF 75lbs / 34kg 375
29 | P a g e PARTS LIST Description Part Number Gasket, Magnetic Door SRI6PF (29 inches X 26 inches) 3450743 Gasket, Magnetic Door SRI20PF (29 inc
3 | P a g e TABLE OF CONTENTS INTRODUCTION ...
Page | 4 INTRODUCTION Thank you for purchasing a Sheldon Refrigerated Peltier Fly Incubator. We know that in today’s competitive marketplace, custo
5 | P a g e INTRODUCTION (CONTINUED) ENGINEERING IMPROVEMENTS Sheldon Manufacturing continually improves all of its products. As a result, enginee
Page | 6 RECEIVING YOUR INCUBATOR Before leaving the factory, all incubators are packaged in high-quality shipping materials to provide protection
7 | P a g e RECEIVING (CONTINUED) RECORDING DATA PLATE INFORMATION Locate the data plate on the back of the incubator at the top right. The data pl
Page | 8 GRAPHIC SYMBOLS The incubator is provided with graphic symbols on its interior and exterior surfaces. The symbols identify hazards and the
9 | P a g e INSTALLATION This incubator is intended for use indoors, at room temperatures between 15C and 30C (59F and 86F), at no greater th