MICROPROCESSOR CONTROLLED OVEN SGO3 SGO3-2 PREVIOUSLY DESIGNATED CE3G CE3G-2 Revised 12/2013 4861567 INSTALLATION AND
10 display digits will light with no lighted decimals showing (See Figure 2). Note that, if during any of the following steps, several seconds elaps
11 MAINTENANCE Warning: Prior to any maintenance or service on this unit, disconnect the power cord from the power source. Before reattaching th
12 TROUBLESHOOTING TEMPERATURE Temperature too high. 1/ Controller set too high-see section 6.3 2/ Controller failed on – call Customer Service. 3
13 1/ ±0.1 may be normal. 2/ For G suffix models: may vary ±2.0 degrees. 3/ For F suffix models: is fan working? –verify movement of air in cham
14 Controller on at all times - "locked-up" 1/ Adjust set point to room temperature. If the unit is still heating, replace the solid st
15 PARTS LIST Description 115v 220V EMI Filter 2800503 2800502 Adjustable Feet 2700512 2700512 Control Knob 4450506 4450506 Cord Set 1800510 1800500
16 UNIT SPECIFICATIONS All Models Weight Shipping Net 200 lbs. 93 lbs. Dimensions Exterior WxDxH (in.) Interior WxDxH (in.) 25.50 x 25.563 x
17 WIRING DIAGRAM SGO3 / SGO3-2 (CE3G / CE3G-2)
2 TABLE OF CONTENTS SECTION 1.0 RECEIVING AND INSPECTION SECTION 2.0 GRAPHIC SYMBOLS SECTION 3.0 INSTALLATION SECTION 4.0 PRECAUTIONS SECTION 5.
3 RECEIVING AND INSPECTION Your satisfaction and safety require a complete understanding of this unit. Read the instructions thoroughly and be sure
4 GRAPHIC SYMBOLS Your oven has been provided with a display of graphic symbols which should help in identifying the use and function of th
5 INSTALLATION Local city, county or other ordinances may govern the use of this equipment. If you have any questions about local requirements, plea
6 PRECAUTIONS 4.1 The bottom surface of the chamber should not be used as a work surface. 4.2 This unit has been designed with a dampered vent from
7 CONTROL PANEL OVERVIEW 5.1 Power Switch: The main power switch on the control panel (green lighted I/O) controls all power to the oven. It must
8 5.7 Temperature Activated Light: This pilot lamp will be lit whenever the elements are receiving power.
9 OPERATION 6.1 Connection to Power Supply: Assure that the electrical power supply is properly configured and rated for the oven and plug the unit